PCR Dichroic Mirrors 62
The beam-splitting interface in a quantitative PCR light path, delivering each channel excitation into the reaction and returning fluorescence to the detector.
View subcategory →PCR fluorescence filters are configured per instrument channel, with at least three elements each: an excitation filter that sets the colour reaching the tube, a dichroic that folds the path, and an emission filter that hands the dye signal cleanly to the detector. Multiplex instruments are the most crosstalk-sensitive of all, so these always ship as matched, pairwise-checked sets against the full wavelength table rather than as loose singles.
| Standard channels | FAM (470/520), HEX (530/560), ROX (580/610), Cy5 (630/680) |
|---|---|
| CWL tolerance | ±1 nm excitation, ±1 nm emission |
| FWHM | Excitation 10–20 nm; emission 20–30 nm |
| Excitation blocking | OD4-OD6 over the specified band; not channel crosstalk |
| Peak transmission | Excitation above 60%; emission above 70% |
| Dichroic steepness | Edge below 5% of the wavelength |
| Size | Standard 25×36 mm or built to the instrument |
| Thermal drift | ≤0.01 nm/℃ over 15–40 ℃ |
Typical processing ranges for this line. Requirements outside these can be evaluated to drawing; final specifications are confirmed by sample and datasheet.
Send the model or drawing — quote and lead time within one business day.
Request a quote